← FDA ENFORCEMENT
Z-1980-2026Class II

Siemens Healthcare Diagnostics, Inc.

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product type
Device
Status
Ongoing
Report date
2026-05-06
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.
Firm state
DE
Product quantity
18,080 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results

Product description

Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

Code / lot info

Material Number: 10872079. Model Number: DF33B. UDI: (01)00630414595009(10)GA6307(17)20261103, (01)00630414595009(10)BA7005(17)20270105. Lot Numbers: GA6307, BA7005.

Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.