← FDA ENFORCEMENT
Z-1894-2026Class II

Magellan Diagnostics, Inc.

Recalling firm
Magellan Diagnostics, Inc.
Product type
Device
Status
Ongoing
Report date
2026-04-29
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bhutan, Bolivia, Canada, Chile, Diego Garcia, Ecuador, Eritrea, Georgia, Ghana, Guam, Guatemala, Haiti, Hong Kong, India, Indonesia, Israel, Italy, Ivory Coast, Japan, Kyrgyzstan, Mexico, Nepal, Peru, Philippines, South Africa, South Korea, Spain, UAE, Uruguay, Vietnam.
Firm state
MA
Product quantity
136
Initiated by
Voluntary: Firm initiated
Reason for recall

Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

Product description

LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)

Code / lot info

Catalog Number: 82-0004; UDI-DI: 00850355006031; ALL lots and ALL sublots;