← FDA ENFORCEMENT
Z-1825-2026Class II

PHILIPS MEDICAL SYSTEMS

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
PHILIPS MEDICAL SYSTEMS
Product type
Device
Status
Ongoing
Report date
2026-04-22
Distribution
Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
Firm state
MA
Product quantity
3 systems
Initiated by
Voluntary: Firm initiated
Reason for recall

Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.

Product description

Philips Spectral CT on Rails. Model Number: 728334.

Code / lot info

Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →