A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Civco Medical Instruments Co. Inc.
Product type
Device
Status
Ongoing
Report date
2026-04-08
Distribution
US States: OH, PA, WA. China.
Firm state
IA
Product quantity
20 units
Initiated by
Voluntary: Firm initiated
Reason for recall

There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

Product description

eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152

Code / lot info

UDI-DI: 00841436111065; Lot Number: A233347, A228677, A268116, A268238, A272513, A279457, A280388 , A281840