Z-1673-2026Class II

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product type
Device
Status
Ongoing
Report date
2026-04-08
Distribution
Domestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, FrenchPolynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Laos, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nige
Product quantity
298 units (27 US, 271 OUS)
Initiated by
Voluntary: Firm initiated
Reason for recall

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

Product description

Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;

Code / lot info

(1) System Code: 722063; UDI: 00884838085275; System Serial Number:122, 56, 64, 131, 84, 132, 96, 98, 108, 77, 82, 145, 47, 146, 80, 102, 43, 117, 2, 149, 99, 44, 3, 28, 144, 62, 119, 61, 22, 29, 153, 40, 70, 30, 25, 19, 120, 121, 110, 14, 16, 17, 53, 103, 127, 97, 143, 41, 104, 116, 151, 93, 141, 115, 137, 73, 150, 128, 105, 4, 50, 129, 72, 38, 21, 78, 23, 76, 63, 69, 59, 51, 92, 139, 130, 154, 75, 24, 94, 100, 54, 46, 26, 71, 106, 83, 36, 37, 27, 15, 135, 123, 107, 87, 85, 89, 88, 86, 90, 91, 79, 133, 81, 140, 52, 48, 148, 55, 113; (2) System Code: 722221; UDI: 00884838085275; System Serial Number:181, 3, 7, 178, 54, 6, 93, 69, 9, 115, 177, 85, 19, 49, 171, 153, 108, 193, 104, 169, 45, 101, 135, 106, 51, 114, 36, 124, 197, 129, 4, 154, 146, 68, 195, 130, 183, 187, 42, 172, 164, 190, 151, 71, 152, 184, 147, 32, 105, 182, 58, 103, 149, 127, 128, 86, 122, 60, 155, 66, 33, 55, 70, 123, 118, 61, 44, 185, 74, 95, 205, 13, 99, 34, 142, 39, 144, 82, 145, 59, 77, 180, 126, 97, 132, 79, 91, 161, 102, 52, 120, 162, 64, 75, 41, 84, 131, 94, 96, 38, 48, 31, 92, 78, 186, 133, 167, 23, 26, 40, 10, 203, 30, 207, 208, 76, 50, 98, 107, 158, 157, 62, 29, 194, 73, 25, 15, 119, 140, 166, 57, 27, 2, 47, 16, 141, 143, 35, 168, 83, 163, 202, 67, 198, 134, 5, 81, 22, 28, 165, 196, 173, 46, 53, 8, 56, 174, 43, 159, 1, 176, 11; (3) System Code: 722229; UDI: 00884838085275; System Serial Number: 22, 9, 40, 7, 24, 35, 16, 15, 41, 5, 3, 29, 4, 38, 33, 32, 6, 2, 20, 26, 31, 10, 11, 21, 25, 13, 8;