Recalling firm
I.T.S. GmbH
Product type
Device
Status
Ongoing
Report date
2026-04-01
Distribution
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Product quantity
11 units
Initiated by
Voluntary: Firm initiated
Reason for recall
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Product description
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Code / lot info
1. Article Number (Lot Number): 70312 (26/F65133). 2. Article Number (Lot Number): 70314 (86/619736).
Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.