I.T.S. GmbH
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.
1. Article Number (Lot Numbers): 21604-6 (643/052301, 643/0652, 643/092201), 21604-9 (643/032311, 643/062308, 643/062309, 643/112310, 643/122201). 2. Article Number (Lot Numbers): 21605-11 (643/072314, 643/102207), 21605-6 (643/0654, 643/072306, 643/072316, 643/072338, 643/102208, 643/102212, 643/102217, 643/112306, 643/122202), 21605-9 (643/0643, 643/0662, 643/072307, 643/072339, 643/112309, 643/122203). 3. Article Number (Lot Numbers): 21105-6 (86/876787), 21105-8 (86/876788). 4. Article Number (Lot Numbers): 21104-4 (679/052234, 679/082305, 86/036334), 21104-5 (679/0722101, 679/082306, 86/036335), 21104-6 (679/012307, 86/876760), 21104-7 (679/082307, 86/036336, 86/876781), 21104-8 (679/082308, 86/876783), 21104-9 (86/876904). 5. Article Number (Lot Numbers): 21101-5 (86/876637), 21101-6 (679/121605, 00/1105-12, 26/F41923, 86/876638, 86/903610), 21101-7 (770/201905, 00/0518-06, 679/042303, 86/876640), 21101-8 (86/876721, 679/062224), 21101-9 (679/122264, 86/876727), 21101-11 (86/876633), 21101-13 (86/876635). 6. Article Number (Lot Numbers): 21102-6 (770/201905). 7. Article Number (Lot Numbers): 21103-8 (679/092216), 21103-9 (679/062215, 86/876758), 21103-10 (679/092219, 86/876741), 21103-11 (679/062214), 21103-12 (86/876743), 21103-13 (679/022303, 86/876745).
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →