I.T.S. GmbH
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article Numbers: 21122-10, 21122-6, 21122-8. 4. Clavicle Plate, Lateral; Article Numbers: 21123-4, 21124-4, 21127-6, 21128-6.
1. Article Number (Lot Numbers): 21116-8 (679/092230), 21117-10 (679/032313, 679/092231), 21118-10 (679/032314, 679/072047, 679/092232). 2. Article Number (Lot Numbers): 21121-6 (679/092236, 679/092329), 21121-8 (679/052226). 3. Article Number (Lot Numbers): 21122-10 (00/0425-02, 00/0825-02, 679/022443, 679/062310, 86/043372), 21122-6 (00/0325-01, 00/0425-01, 679/042340, 679/062334, 679/122269, 86/043366), 21122-8 (00/0325-08, 679/022243, 679/022304, 679/022451, 679/042209, 679/042214, 679/052319, 679/102368, 86/043368). 4. Article Number (Lot Numbers): 21123-4 (00/0810-06, 679/052014, 679/062216), 21124-4 (679/042333, 679/062217, 679/072346), 21127-6 (679/052228), 21128-6 (679/042341).
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →