Z-1488-2026Class II

Olympus Corporation of the Americas

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Olympus Corporation of the Americas
Product type
Device
Status
Ongoing
Report date
2026-03-11
Distribution
US Nationwide distribution.
Firm state
PA
Product quantity
95,882 boxes (20 per box)
Initiated by
Voluntary: Firm initiated
Reason for recall

Potential for rubber fragment detachment during use.

Product description

Olympus Single use Biopsy Valve. Model/Catalog Number: MAJ-210. 20 units per box. The MAJ-210 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent reflux of body fluids.

Code / lot info

Model/Catalog Number: MAJ-210. UDI-DI: 14953170152433, 14953170452069. All Lot Numbers. 20 units per box.