Staar Surgical AG
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →