← FDA ENFORCEMENT
Z-1383-2026Class II

Auris Health, Inc

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Auris Health, Inc
Product type
Device
Status
Ongoing
Report date
2026-02-25
Distribution
Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
Firm state
CA
Product quantity
173
Initiated by
Voluntary: Firm initiated
Reason for recall

Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.

Product description

MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software

Code / lot info

Software versions: 2.2.2, 2.2.3, 2.2.4, 2.2.5, and 4.1.1. UDI-DI: 10810068810803, 10810068810988. Serial Numbers: 110024, 120088, 120057, 120046, 120058, 120015, 120160, 120068, 120081, 120056, 110028, 120128, 120017, 120089, 120075, 110016, 120031, 120169, 120113, 120168, 120055, 120014, 120060, 120174, 120195, 120026, 110038, 120124, 120194, 110027, 110035, 120123, 120091, 120007, 110022, 120006, 120152, 120004, 120120, 120041, 120106, 110034, 120126, 120082, 120151, 120051, 120020, 120153, 120193, 120170, 120108, 120037, 120074, 120199, 120130, 120095, 110026, 120096, 120009, 120111, 120022, 120054, 120163, 110045, 120027, 110039, 120090, 120141, 120072, 120155, 110049, 120143, 110047, 120125, 120052, 110052, 120100, 120050, 120098, 120045, 120131, 110020, 110030, 110032, 110021, 120013, 120105, 120136, 120192, 110017, 120029, 120139, 110031, 120137, 120053, 110029, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 110033, 120145, 120127, 110046, 110043, 120011, 120070, 120076, 120134, 120117, 110025, 120077, 120149, 120191, 120102, 120002, 120156, 120010, 120171, 120003, 120023, 120146, 120188, 120043, 120071, 120042, 120196, 110051, 120065, 120172, 120116, 120150, 120066, 120166, 120033, 110013, 120079, 120008, 120086, 120032, 120073, 1806P1, 120197, 120080, 120078, 14, 120012, 110041, 120084, 120219, 110036, 120181, 120083, 120067, 120021, 120216, 120217, 120218, 120069, 120204, 120207, 120215, 120237, 120208, 120213, 120164, 120132, 120182

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →