Fujirebio Diagnostics, Inc.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio is an in vitro diagnostic (IVD) test that combines the test results of the Lumipulse G pTau 217 Plasma assay and the Lumipulse G ¿-Amyloid 1-42-N Plasma assay from the same patient specimen (K2EDTA plasma sample) into a numerical ratio. The Lumipulse G pTau 217/¿-Amyloid 1-42 Plasma Ratio Instructions For Use (IFU), CL0072, contains instructions on how to manually calculate the ratio and is supplied with each individual Lumipulse G assay. Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio, Item# CL0073 is comprised of the following kits, which are imported: Kit Description, Item#: Imported from Fujirebio Europe (FRE): Lumipulse G pTau 217 Plasma Immunoreaction Cartridges 81550 Lumipulse G pTau 217 Plasma Calibrators set 81557 Lumipulse pTau 217 Plasma Controls 81558 Lumipulse ¿-Amyloid Plasma Controls 81559 Imported from Fujirebio Japan (FRI): Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges 235447 Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators set 235454 Component: No
Model/Catalog Number: CL0073; UDI-DI: 00869164000086; Distribution for Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio ("Ratio") is related to the "Ratio" Instructions for Use (IFU). Therefore, each consignee has 1 "Ratio" in which they purchase the component kits for.
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →