← FDA ENFORCEMENT
Z-1237-2026Class II

Tornier, Inc

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Tornier, Inc
Product type
Device
Status
Ongoing
Report date
2026-02-11
Distribution
US Nationwide distribution.
Firm state
MN
Product quantity
427
Initiated by
Voluntary: Firm initiated
Reason for recall

A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.

Product description

TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318

Code / lot info

Lot AZ4824290

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →