Beckman Coulter, Inc.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
CellMek SPS Sample Preparation System, REF: C44603, with software
UDI-DI: 15099590750312. Software Version 2.3.90.0