← FDA ENFORCEMENT
Z-1162-2026Class I

Boston Scientific Corporation

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Boston Scientific Corporation
Product type
Device
Status
Ongoing
Report date
2026-02-11
Distribution
Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.
Firm state
MA
Product quantity
150 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

Product description

*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;

Code / lot info

Material Number (UPN): M00553530; UDI-DI: 8714729904557; Lot Numbers: 35985705, 35986358, 36147098, 36147222, 36165642, 36176423, 36190575, 36191222, 36194448, 36218745, 36227778, 36227779, 36241500, 36271259, 36280398, 36293880, 36300839, 36301620, 36343103, 36343104, 36364646, 36425887, 36469811, 36471228, 36491879, 36492660, 36506029, 36510220, 36510221, 36513401, 36513402, 36513403, 36517139, 36517220, 36525208, 36525209, 36525210, 36534514, 36534518, 36536259, 36536480, 36536481, 36551000, 36551001, 36642564, 36642565, 36642756, 36645775, 36663654, 36677842, 36677843, 36682759, 36684887, 36684888, 36730665, 36730666, 36747824, 36747825, 36753898, 36753899, 36757493, 36757494, 36761827, 36766370, 36766371, 36783265, 36792467, 36792468, 36792469, 36792470, 36792471, 36799315, 36815090, 36815092, 36832774, 36833683, 36836357, 36844783, 36845422, 36846169, 36846176, 36851617, 36851792, 36853596, 36865505, 36873631, 36873632, 36873906, 36874863, 36904228, 36904229, 36904230, 36907002, 36907003, 36907004, 36907028, 37102533, 37102534, 37102535, 37102536, 37102537, 37459567, 37459568, 37459569, 37459570, 37463133, 37470670, 37498192, 37511243, 37519095, 37519096, 37844029, 37519097, 37520927, 37530320, 37530321, 37530333, 37533308, 37541933, 37571927, 37571928, 37579364, 37579366, 37579367, 37580630, 37585525, 37585526, 37589261, 37589262, 37594668, 37649481, 37649482, 37649483, 37650631, 37694052, 37699011, 37699012, 37708202, 37779125, 37779126, 37780822, 37780823, 37780824, 37823589, 37823590, 37827568, 37827569, 37833165, 37833176, 37834499;

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →