← FDA ENFORCEMENT
Z-1120-2026Class II

ICU Medical, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
ICU Medical, Inc.
Product type
Device
Status
Ongoing
Report date
2026-01-28
Distribution
Worldwide distribution - US Nationwide and the country of Philippines.
Firm state
IL
Initiated by
Voluntary: Firm initiated
Reason for recall

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Product description

ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Code / lot info

400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →