RAYSEARCH LABORATORIES AB
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
Lot Code: UDI: 0735000201082220250423, 0735000201094520250702 GTIN: 07350002010822, 07350002010945 Serial Numbers: 17.0.0.1270, 17.0.1.113 Software Revision: RayStation v2025, RayStation v2025 SP1 Expiration Date: 2030-10-22