← FDA ENFORCEMENT
Z-1105-2026Class II

RAYSEARCH LABORATORIES AB

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
RAYSEARCH LABORATORIES AB
Product type
Device
Status
Ongoing
Report date
2026-01-28
Distribution
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Product quantity
8
Initiated by
Voluntary: Firm initiated
Reason for recall

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Product description

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Version: RayStation 12A, RayStation 12A SP1, RayStation 12A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No

Code / lot info

Lot Code: UDI: 0735000201054920220616, 0735000201067920221007, 0735000201073020230913 GTIN: 07350002010549, 07350002010679, 07350002010730 Serial Numbers: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Revisions: RayStation 12A, 12A SP 1, 12A SP2 Expiration Date: 2028-07-07

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →