RAYSEARCH LABORATORIES AB
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System
UDI: 0735000201042620211208, 0735000201049520220312, 0735000201050120220422, 0735000201060020220620 and 0735000201057020221222. GTIN: 07350002010426, 07350002010495, 07350002010501, 07350002010600 and 07350002010570. Serial Numbers: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12, 12.3.0.119. Software Revisions: RayStation 11B, 11B SP1, 11B SP2, 11B SP3, 11B SPT1. Expiration Date: 2028-03-20.