← FDA ENFORCEMENT
Z-1051-2026Class II

Medline Industries, LP

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2026-01-21
Distribution
US Nationwide distribution.
Firm state
IL
Product quantity
32 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.

Product description

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DELUXE PRE OP KIT, Medline SKU DYKS2034A; 2) DELUXE PRE OP KIT, Medline SKU DYKS2006B.

Code / lot info

Medline kit SKU PHS972096014B UDI/DI: 10889942695911 (EA) 40889942695912 (CS), Lot Number 25IBT843; Medline kit SKU CDS984289O UDI/DI: 10198459418143 (EA) 40198459418144 (CS), Lot Number 25IMA327; Medline kit SKU DYNJ903523F UDI/DI: 10198459316593 (EA) 40198459316594 (CS), Lot Number 25IMI028.

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →