← FDA ENFORCEMENT
Z-1037-2026Class II

Integra LifeSciences Corp. (NeuroSciences)

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Integra LifeSciences Corp. (NeuroSciences)
Product type
Device
Status
Ongoing
Report date
2026-01-14
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
Firm state
NJ
Product quantity
5,116 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.

Product description

Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a 14-gauge Tuohy needle with stylet. The ICP Sensor is a nylon tube with a microminiature strain gauge pressure transducer (sensing element) mounted at one end and an electrical connector at the other end. It is designed for use with a Codman intracranial pressure monitoring device. The ICP Sensor Basic Kit is designed for use with the Codman Cranial Hand Drill. The drill facilitates access to the intraparenchymal area. The drill is also available as a component of the Codman Cranial Access Kit. Use of the ICP Sensor Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.

Code / lot info

Catalog Number: 826850. UDI-DI: 10381780520672. Lot Numbers: 7261958, 7266691, 7266692, 7288060, 7293837, 7293836, 7306040, 7306041, 7306042, 7306043, 7306044, 7306045, 7306046, 7306047, 7306050, 7306051, 7306052, 7307670, 7307671, 7307672, 7325482, 7307675, 7325481, 7325483, 7325484, 7307673, 7307674, 7325485, 7325486, 7336611, 7336612, 7336614, 7336623, 7336613, 7336615, 7336616, 7336618, 7336619, 7336620, 7336621, 7336622, 7306048, 7306049, 7336626, 7336627, 7336629, 7347166, 7336628, 7346295, 7346297, 7346299, 7346301, 7365434, 7365435, 7392019, 7396607, 7396608, 7396609, 7396610, 7392034, 7392067, 7405790, 7405791, 7405792, 7405793, 7406633, 7406634, 7406636, 7406637, 7406639, 7406640, 7406641, 7406642, 7406635, 7406638, 7408813, 7408815, 7408818, 7408814, 7408816, 7468726, 7468727, 7411801, 7411802, 7414641, 7414642, 7468731, 7468732, 7468745, 7468746, 7468741, 7468742, 7468743, 7468744, 7468771, 7468769, 7468770, 7468773, 7468774, 7468775, 7482179, 7482180, 7482181, 7482182, 7482183, 7482184, 7482185, 7482186, 7468772, 7482154, 7482155, 7482156, 7482157, 7468801, 7468804, 7468805, 7468806, 7482136, 7468802, 7468803, 7482132, 7482134, 7482135, 7482137, 7482159, 7482158, 7482160, 7482161, 7495593, 7495594, 7495596, 7495597, 7495598, 7495599, 7482212, 7495600, 7495601, 7495602, 7495603, 7495604, 7495605, 7495606, 7495607, 7495608, 7495635, 7495636, 7495637, 7495638, 7495639, 7495640, 7495641, 7495642, 7495643, 7495666, 7495644, 7495664, 7482213, 7495667, 7503597, 7503598, 7503599, 7503659, 7510140, 7510141, 7510142, 7503600, 7503601, 7517716, 7517708, 7503627, 7503628, 7503630, 7503631, 7503632, 7503633, 7503629, 7503664, 7503657, 7503662, 7503663, 7503602, 7503624, 7510156, 7517703, 7517704, 7517706, 7517707, 7517709, 7517710, 7517711, 7517712, 7517714, 7517715, 7408817, 7411797, 7411798, 7411799, 7411800, 7414643, 7414644, 7414645, 7414646, 7414647, 7414648, 7414649, 7417938, 7417940, 7417942, 7417944, 7417946, 7417948, 7417950, 7417952, 7417955, 7417957, 7417959, 7503625.

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →