← FDA ENFORCEMENT
Z-1024-2026Class II

REPLIGEN CORPORATION

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
REPLIGEN CORPORATION
Product type
Device
Status
Ongoing
Report date
2026-01-21
Distribution
U.S.
Firm state
MA
Product quantity
33
Initiated by
FDA Mandated
Reason for recall

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Product description

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Code / lot info

PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496