← FDA ENFORCEMENT
Z-1015-2026Class I

Medline Industries, LP

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2026-01-28
Distribution
US, CA, PA, GUAM, VI
Firm state
IL
Product quantity
35694 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Product description

MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L

Code / lot info

UDI/DI 40080196294344, All lots

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →