Vision RT Ltd

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Vision RT Ltd
Product type
Device
Status
Ongoing
Report date
2026-01-14
Distribution
U.S. and OUS
Product quantity
56
Initiated by
FDA Mandated
Reason for recall

Vision RT is informing customers of an omission identified in the Instructions for Use for AlignRT InBore systems. Information identifying the lasers and their characteristics was omitted from the Instructions for Use for the subject system as required by 21 CFR 1040,10(h), via conformance with recognized consensus standard IEC 60825-1 Ed. 3 (FDA Laser Notice 56).

Product description

The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.

Code / lot info

Part Number V000629