Z-0965-2026Class II

ICU Medical, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
ICU Medical, Inc.
Product type
Device
Status
Ongoing
Report date
2025-12-31
Distribution
US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.
Firm state
IL
Product quantity
7 customers
Initiated by
Voluntary: Firm initiated
Reason for recall

For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.

Product description

LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02

Code / lot info

version 2.2.1 UDI-DI 00887709127227 Issue affects customers who migrated from LifeShield DLM v2.1x to LifeShield DLM v2.2.1 and have any version of a Plum Duo v1.1 drug library that was approved or locked for review before the migration.