← FDA ENFORCEMENT
Z-0963-2026Class II

Howmedica Osteonics Corp.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Howmedica Osteonics Corp.
Product type
Device
Status
Ongoing
Report date
2025-12-31
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.
Firm state
NJ
Product quantity
320 units
Initiated by
Voluntary: Firm initiated
Reason for recall

When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

Product description

The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.

Code / lot info

Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;