← FDA ENFORCEMENT
Z-0952-2026Class II

Rocket Medical Plc

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Rocket Medical Plc
Product type
Device
Status
Ongoing
Report date
2025-12-24
Distribution
Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Product quantity
630 units
Initiated by
Voluntary: Firm initiated
Reason for recall

IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.

Product description

Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54544-18-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A

Code / lot info

Model/Catalog Number: R54544-18-PK; UDI-DI: 050552709TF06DXY; Lot Code: 496899, 500956, 497076, 501169, 497584, 501188, 499227, 502430, 499417, 502951, 499461, 503306, 499897, 503430, 499898, 503529, 499899, 504180, 499908, 504543, 500407, 504823, 500477;

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →