← FDA ENFORCEMENT
Z-0923-2026Class II

AGFA Healthcare Corp.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
AGFA Healthcare Corp.
Product type
Device
Status
Ongoing
Report date
2025-12-31
Distribution
U.S. Nationwide distribution.
Firm state
SC
Product quantity
35
Initiated by
FDA Mandated
Reason for recall

It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

Product description

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Code / lot info

N/A

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →