← FDA ENFORCEMENT
Z-0890-2026Class II

Waldemar Link GmbH & Co. KG (Mfg Site)

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product type
Device
Status
Ongoing
Report date
2025-12-17
Distribution
US distribution to states of: AL, GA, and TX.
Firm state
N/A
Product quantity
18 units
Initiated by
Voluntary: Firm initiated
Reason for recall

The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.

Product description

Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.

Code / lot info

Item Number: 15-0028/08; UDI-DI: 04026575175222; Lot Number: 2535004;

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →