← FDA ENFORCEMENT
Z-0600-2026Class II

Stryker Corporation

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Stryker Corporation
Product type
Device
Status
Ongoing
Report date
2025-12-03
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
Firm state
MI
Product quantity
39,148
Initiated by
Voluntary: Firm initiated
Reason for recall

Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.

Product description

NICO Myriad Illumination Pack Product No. NN-1000 11g Illumination Sleeve/Fiber; Product No. NN-1001 13g Illumination Sleeve/Fiber

Code / lot info

Product Number/UDI-DI (GTIN)/Lot Numbers: NN-1000 00816744021009 37AC5163 37AE9811 37AJ1438 37BJ0513 37CB3410 37CB3416 37CD6003 37CE3030 37EB5720 37FA4260 37GD2216 37GJ1603 37IC7023 37IJ0208 37JJ0122 37KI1300 37LA0158 37LI0320 37LJ1522; NN-1001 00816744021016 37AC5177 37AE9942 37AE9955 37AJ1439 37BJ0516 37CB3380 37CB3419 37CD6005 37CJ0204 37DE4864 37EB5721 37FA4261 37GD2219 37GJ1607 37IC7027 37IJ0209 37JD5753 37JJ0123 37KI1301 37LA0159 37LI0322 37LJ1525

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →