Magellan Diagnostics, Inc.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare¿ II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Product Name: LeadCare¿ II Blood Lead Test Kit; Catalog Number: 70-6762; Impacted Lot/Sublot Numbers: 2333M, 2333M-01, 2333M-02, 2333M-03, 2333M-04, 2333M-05, 2333M-06, 2333M-07, 2333M-08, 2333M-09, 2333M-10, 2333M-11; Device Identifier: (01)00850355006000;