← FDA ENFORCEMENT
Z-0584-2026Class II

Magellan Diagnostics, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Magellan Diagnostics, Inc.
Product type
Device
Status
Ongoing
Report date
2025-12-03
Distribution
US Nationwide distribution.
Firm state
MA
Product quantity
1,774 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.

Product description

The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare¿ II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)

Code / lot info

Product Name: LeadCare¿ II Blood Lead Test Kit; Catalog Number: 70-6762; Impacted Lot/Sublot Numbers: 2333M, 2333M-01, 2333M-02, 2333M-03, 2333M-04, 2333M-05, 2333M-06, 2333M-07, 2333M-08, 2333M-09, 2333M-10, 2333M-11; Device Identifier: (01)00850355006000;