← FDA ENFORCEMENT
Z-0542-2026Class II

Cellavision AB

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Cellavision AB
Product type
Device
Status
Ongoing
Report date
2025-11-26
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.
Product quantity
688 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.

Product description

REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.

Code / lot info

UDI: 0735004097202/ Serial Numbers 63220-70093 or CN63220-CN70093 Serial Numbers 60001-63219 or CN60001-CN63219