← FDA ENFORCEMENT
Z-0522-2026Class II

Medline Industries, LP

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2025-11-26
Distribution
US Nationwide distribution.
Firm state
IL
Product quantity
1315 kits
Initiated by
Voluntary: Firm initiated
Reason for recall

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Product description

MEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202R; 2) ACDF CDS, REF CDS941061M; 3) NEURO PACK-LF, REF DYNJ0038440C; 4) NEURO VP SHUNT/VAGAL NERVE-RF, REF DYNJ21925S; 5) NEURO-ORTHO EXTRAS, REF DYNJ44882M; 6) LAMINECTOMY, REF DYNJ903870F.

Code / lot info

1) REF CDS780202R, UDI/DI 10195327418618 (each) 40195327418619 (case), Lot Numbers: 23LMG063; 2) REF CDS941061M, UDI/DI 10193489445589 (each) 40193489445580 (case), Lot Numbers: 24ABG356, 24CBH117, 24DBJ347; 3) REF DYNJ0038440C, UDI/DI 10195327356699 (each) 40195327356690 (case), Lot Numbers: 23JMC956, 23KMF585, 24AMB974; 4) REF DYNJ21925S, UDI/DI 10195327105020 (each) 40195327105021 (case), Lot Numbers: 23EBS176, 23GBF899, 23HBZ085; 5) REF DYNJ44882M, UDI/DI 10195327382803 (each) 40195327382804 (case), Lot Numbers: 23FDA522, 23GDB903; 6) REF DYNJ903870F, UDI/DI 10195327239268 (each) 40195327239269 (case), Lot Numbers: 23IBD913, 23IBD914, 23JBL889, 24ABE596, 24BBL042, 24CBH196, 24DBU825.