← FDA ENFORCEMENT
Z-0503-2026Class II

ELEKTA SOLUTIONS AB

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
ELEKTA SOLUTIONS AB
Product type
Device
Status
Ongoing
Report date
2025-11-26
Distribution
Worldwide - U.S. Nationwide distribution in the state of NM and the countries of Canada, Malta, and South Africa.
Product quantity
4 systems that contain the affected software.
Initiated by
Voluntary: Firm initiated
Reason for recall

When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.

Product description

MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0 MOSAIQ¿ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.

Code / lot info

Software Versions: 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 UDI codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →