← FDA ENFORCEMENT
Z-0501-2026Class II

TECHNO-PATH MANUFACTURING LTD.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
TECHNO-PATH MANUFACTURING LTD.
Product type
Device
Status
Ongoing
Report date
2025-11-19
Distribution
US distribution to CA, CO, CT, FL, HI, IL, LA, MI, MN, NV, NY, OR, SC, TN, TX, WI, WY.
Product quantity
76 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Potential of negative recovery of the HBsAg analyte across two (2) lots of Multichem ID-B Positive Control when tested on the Alinity i platform.

Product description

Multichem ID-B. Model/Catalog Number: SR103B Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory testing procedures for the determination of anti-HIV-1 IgG, anti-HBc IgG, anti-HTLV-I IgG, anti-HCV IgG, HBs antigen and anti-Treponema pallidum IgG.

Code / lot info

Model Number: SR103B. UDI: (01)05391523442312(240)SR103B(17)260531(10)SB140524P. Lot number: SB140524P. Expiration Date: 31-5-2026