Siemens Healthcare Diagnostics, Inc.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Material Number (SMN): 10385891; Model/Catalog: FP3L4; UDI-DI: 00630414960418; Lot number: 629;