← FDA ENFORCEMENT
Z-0452-2026Class II

Olympus Corporation of the Americas

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Olympus Corporation of the Americas
Product type
Device
Status
Ongoing
Report date
2025-11-19
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, India, Taiwan, Germany, Australia, Singapore.
Firm state
PA
Product quantity
1528 units
Initiated by
Voluntary: Firm initiated
Reason for recall

The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.

Product description

Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones. Component: ShockPulse Lithotripsy System

Code / lot info

Model/Catalog Number: SPL-T; UDI-PI: 00855279005023; All serial numbers;