← FDA ENFORCEMENT
Z-0449-2026Class I

Medline Industries, LP

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2025-11-19
Distribution
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Firm state
IL
Product quantity
96,406 total units
Initiated by
Voluntary: Firm initiated
Reason for recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Product description

Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230

Code / lot info

CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230 UDI-DI 40884389688589 (case), 10884389688588 (ea) Lots 25EBV570 25FBT843

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →