Spacelabs Healthcare, Ltd.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.
Catalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →