← FDA ENFORCEMENT
Z-0322-2026Class II

Surgical Theater Inc

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Surgical Theater Inc
Product type
Device
Status
Terminated
Report date
2025-11-05
Distribution
Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Firm state
OH
Product quantity
14 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Data Integrity
Reason for recall

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Product description

SyncAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SNP.1.12.0, imaging processing radiological system

Code / lot info

UDI/DI 00860003884102

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →