← FDA ENFORCEMENT
Z-0303-2026Class II

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Product type
Device
Status
Ongoing
Report date
2025-10-29
Distribution
Worldwide
Product quantity
102 units
Initiated by
Voluntary: Firm initiated
Reason for recall

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Product description

Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Code / lot info

all serial numbers in distribution

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →