Recalling firm
Philips North America
Product type
Device
Status
Ongoing
Report date
2025-10-22
Distribution
US Nationwide distribution in the states of CA, IA, TN, TX, VA.
Firm state
MA
Product quantity
28 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.

Product description

Cardiac Workstation 5000; Model Number: 860439;

Code / lot info

Model Number: 860439; UDI-DI: 00884838094826; Serial Numbers: