← FDA ENFORCEMENT
Z-0155-2027Class II

Steris Corporation

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Steris Corporation
Product type
Device
Status
Ongoing
Report date
2026-10-21
Distribution
Worldwide - US Nationwide distribution.
Firm state
OH
Product quantity
11806 units
Initiated by
Voluntary: Firm initiated
Reason for recall

If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.

Product description

RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116

Code / lot info

UDI: 00724995245955 Lot numbers with expiration dates: 689171 5/4/2027 689981 5/12/2027 690993 5/26/2027 691739 6/2/2027 692570 6/11/2027 694357 7/1/2027

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →