← FDA ENFORCEMENT
Z-0145-2026Class II

Kico Knee Innovation Company

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Kico Knee Innovation Company
Product type
Device
Status
Ongoing
Report date
2025-10-22
Distribution
US Nationwide distribution in the states of IN, MN, MO, OH, FL, IL, RI, KY, CA, WI, MI, PA, WA, CO, AL.
Product quantity
21 systems
Initiated by
Voluntary: Firm initiated
Reason for recall

Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.

Product description

Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version: V2025.1.2 Product Description: Intended Use: The ARVIS" system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. Description: ARVIS Shoulder is a computer-controlled surgical navigation system for shoulder arthroplasty. It aids the surgeon in making intra-operative measurements and locating anatomical structures of the shoulder joint based on the patient s preoperative imaging to assist with selection and positioning of orthopedic implant components. The system consists of software, electronic hardware and surgical instruments. The device s workflow involves CT based preoperative planning followed by intraoperative navigation and execution. The preoperative planning software enables 3D virtual implant positioning based on the patient s CT reconstructed digital bone model and bony landmarks. The shoulder navigation application software then matches the patient s digital bone model and landmarks to the intraoperative landmarks registered by the surgeon, allowing an image-based navigation to follow. The surgeon uses the plan data as guidance to navigate and help position shoulder instruments and implants.

Code / lot info

Software Version V2025.1.2

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →