← FDA ENFORCEMENT
Z-0132-2026Class II

Insulet Corporation

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Insulet Corporation
Product type
Device
Status
Ongoing
Report date
2025-10-22
Distribution
US Nationwide distribution.
Firm state
MA
Product quantity
174,013 units
Initiated by
Voluntary: Firm initiated
Reason for recall

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Product description

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

Code / lot info

Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X