← FDA ENFORCEMENT
Z-0108-2026Class II

MICROVENTION INC.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
MICROVENTION INC.
Product type
Device
Status
Ongoing
Report date
2025-10-22
Distribution
Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY.
Firm state
CA
Product quantity
467
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.

Product description

Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X

Code / lot info

REF/UDI-DI(Lot): FRED5019/00811425020388(0000235839), 00842429117163(0000544498); FRED5029/00811425020395(0000256343), 00842429117170(0000507209); FRED4528/00811425020425(0000265666); FRED4026/00811425021033(0000333125, 0000576973); FRED5526/00842429117194(0000361117); FRED4508/00842429117095(0000372088); FRED5514/00842429117187(0000429786, 0000478385); FRED3536/00842429117033(0000460268, 0000445555); FRED3511/00842429117002(0000452923); FRED5014/00842429117156(0000563591, 0000464401); FRED3516/00842429117019(0000513146, 0000531733, 0000536217); MV-F451827/04987892121906(0000565094); XFRED4017-MVE/00842429107508(0000478838); XFRED4528-MVE/00842429107560(0000518323, 0000663751); MV-F501427X/04987892151231(0000579031); XFRED4518-MVE/00842429107553(0000663074, 0000663750); XFRED5526-MVE/00842429107638(0000704203); XFRED4539-MVE/00842429107577(00007311620); FREDX5522-PMA/00842429114483(0000265641, 0000469609, 0000731155); FREDX3522-PMA/00842429114315(0000469598); FREDX4525-PMA/00842429114414(0000469603); FREDX5015-PMA/00842429114445(0000469606); FREDX5532-PMA/00842429114490(0000469610); FREDX4013-PMA/00842429114346(0000469613, 0000663758, 0000668189); FREDX3513-PMA/00842429114292(0000478834); FREDX4515-PMA/00842429114391(0000478840); FREDX3517-PMA/00842429114308(0000731304); MV-F352427X/04987892151101(0000760963)

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →