Z-0028-2026Class I

NOXBOX LTD

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
NOXBOX LTD
Product type
Device
Status
Ongoing
Report date
2025-10-15
Distribution
Domestic: TN
Product quantity
1667 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

Product description

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Code / lot info

Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.