· EQUIPMENT & PROCESS LIBRARY

VHP bio-decontamination system

Vaporised hydrogen peroxide generator · VHP · VPHP · H2O2 decontamination

E09 · ISOLATORS, RABS AND CONTAINMENTAseptic processingChecked against primary sources by SPEQ

The system that takes an isolator interior through an automated sporicidal gassing cycle: its phases, what it measures, and why a concentration reading is never proof of kill.

What this page does not claim

An equipment family page teaches how a kind of equipment works and how it is assured. It is not a qualification of any site’s equipment, sets no universal limits, and was checked by SPEQ against the sources it lists, not by an independent engineer.

WHAT IT DOES

Annex 1 defines bio-decontamination as "a process that eliminates viable bioburden via use of sporicidal chemical agents". For isolators it asks that the process "should be automated, validated and controlled within defined cycle parameters and should include a sporicidal agent in a suitable form (e.g. gaseous or vaporized form)" (4.22 i).

A vaporised hydrogen peroxide (VHP) system does that: it conditions the enclosure air, introduces hydrogen peroxide vapour, holds the validated conditions, then removes the peroxide so people and product are not exposed to it. Cleaning comes first: residues left on surfaces "may inhibit the effectiveness of the decontamination process" (4.22).

Where it sits: Before an aseptic campaign, after cleaning and set-up, and after any opening of the isolator that breaks its isolation; also in material airlocks and transfer chambers.

WHERE YOU FIND IT

  • Isolators for aseptic filling and compounding
  • Sterility-testing isolators
  • Material transfer airlocks
  • Room or enclosure decontamination where validated

In the reference facility: Aseptic filling & isolator line, Microbiology & environmental monitoring laboratory

Not covered on this page: Sterilisation of product-contact parts, which are sterilised by a validated sterilisation process; Routine sporicidal disinfection of RABS by manual application.

MAJOR COMPONENTS

  • Generator (vaporiser and injection). Turns liquid hydrogen peroxide into vapour and doses it into the enclosure.
  • Air handling and dehumidification. Conditions the enclosure air before gassing and circulates the vapour.
  • Distribution (nozzles, fans, circulation path). Carries vapour to every interior surface, including gloves extended with fingers separated (Annex 1, 4.22 i).
  • Catalytic converter or extraction. Breaks down or removes peroxide during aeration.
  • Sensors and cycle controller. Run the phases against defined parameters and record them.

Connects to: The isolator enclosure and its doors, gloves and transfer ports; Building exhaust for aeration; Hydrogen peroxide supply and its storage; Batch or cycle records.

INSTRUMENTS AND WHAT THEY MEASURE

  • Hydrogen peroxide concentration sensor. Peroxide concentration during gassing and dwell, and the low residual after aeration. Watch for: a reading at one point taken to represent every surface.
  • Humidity sensor. Humidity during conditioning and gassing, which affects condensation and distribution. Watch for: condensation or peroxide exposure shifting the sensor.
  • Temperature sensor (RTD or thermocouple). Enclosure and surface temperature, which drive condensation. Watch for: sensor not at the location that matters (cold spot, load centre).
  • Differential pressure transmitter. Enclosure pressure during the cycle, so vapour does not escape and room air does not enter. Watch for: ports connected the wrong way round, reversing the sign.

PARAMETERS

SETPOINT · MEASURED · LIMIT · ATTRIBUTE
  • Relative humidity (setpoint, % RH). The conditioning target before gassing, defined in cycle development for this enclosure and load.
  • Hydrogen peroxide concentration (measured, ppm). At the sensor location during gassing and dwell; one point, not every surface.
  • Exposure (dwell) time (setpoint, min). The dwell time proven in cycle validation for the defined load configuration.
  • Residual hydrogen peroxide (limit, ppm). After aeration, before doors open or product is exposed; the site justifies the limit for people and product separately.
  • Temperature (measured, °C). Enclosure air and the coldest surfaces, where condensation starts.

SPEQ publishes no universal alarm or acceptance values: they depend on the process, the product and the site’s own qualification.

CONTROLS, ALARMS AND INTERLOCKS

  • Control loop: Cycle phase control. Moves through conditioning, gassing, dwell and aeration on defined parameters.
  • Alarm: Out-of-range cycle parameter. Flags a cycle that did not run within its validated parameters so it is not released as valid.
  • Interlock: Door lock until residual is below limit. Keeps the isolator closed until aeration has brought peroxide below the defined level.
  • Safety function: Room peroxide detection. Protects people outside the enclosure from leaks; a safety requirement with its own owner, separate from product quality.

An alarm asks for a response; an interlock stops an action; a safety function protects people and has its own risk basis.

HOW IT FAILS

  • A surface the vapour does not reach (shadowed or occluded). Viable organisms survive where the indicator was not placed.
  • Uncontrolled condensation. Uneven exposure and possible material damage or residue.
  • Load or set-up different from the validated configuration. The validated cycle may no longer reach every surface.
  • Incomplete aeration. Residual peroxide can affect product or personnel.
  • Cleaning skipped or poor. Residues inhibit decontamination (Annex 1, 4.22).

ASSURANCE ACROSS THE LIFECYCLE

EXAMPLES, NOT A MANDATORY LIST
  • Cycle development. Mapping of distribution and the hardest-to-reach locations; Defined load configurations; Material compatibility, including evidence agents do not adversely affect product (Annex 1, 4.22).
  • Cycle validation. Biological indicators at worst-case locations; Geobacillus stearothermophilus spores on stainless steel are the routine indicator for hydrogen peroxide processes; Chemical indicators for distribution; Aeration studies to the residual limit.
  • Routine operation. Each cycle’s parameter record reviewed against its validated range; Periodic indicator challenge; Glove positioning checks before gassing.
  • Change and requalification. Impact assessment of any change to load, fixtures, gloves or generator; Requalification where distribution could change.

WHAT THIS DOES NOT PROVE

  • A peroxide concentration reading does not prove biological decontamination. Instead: Biological indicators at the worst-case locations, in the validated load.
  • A successful cycle does not prove suitability for every load or configuration. Instead: Validation of each load configuration the site uses.
  • A residual below the personnel limit does not prove there is no effect on the product. Instead: Product-impact evidence for the agents used (Annex 1, 4.22).

SOURCES

4 CITED

HOW THIS PAGE WAS CHECKED

Researched and checked against the sources above by SPEQ (Claude) on 2026-10-21. It has not been reviewed by an independent engineer. It teaches how this kind of equipment works and is assured; it does not say any site, asset or process is qualified, validated or compliant.

Spotted something wrong? Tell us.