E33 · CONTROLLED-TEMPERATURE AND CRYOGENIC EQUIPMENTControlled storageChecked against primary sources by SPEQ

The chambers and rooms that hold materials, product and stability samples at controlled temperature and humidity, and the mapping, monitoring and alarms that show they did.

What this page does not claim

An equipment family page teaches how a kind of equipment works and how it is assured. It is not a qualification of any site’s equipment, sets no universal limits, and was checked by SPEQ against the sources it lists, not by an independent engineer.

WHAT IT DOES

Controlled-temperature equipment keeps materials and product within their labelled storage conditions and holds stability samples at the conditions a study requires. Chapter 3 asks that where special storage conditions are required they be "provided, checked and monitored" (3.19).

Stability chambers support the on-going stability programme, which monitors marketed product so stability issues are detected; Chapter 6 asks that the equipment used, "stability chambers among others", be qualified and maintained following Chapter 3 and Annex 15 (6.26, 6.29).

Where it sits: In the warehouse for materials and finished product, in production for intermediates, and in the laboratory for samples and stability studies.

WHERE YOU FIND IT

  • Cold storage of biologics, vaccines and temperature-sensitive materials
  • Stability studies at defined temperature and humidity
  • Storage of reference standards, reagents and media

In the reference facility: Stability storage & sample management, Quarantine & released warehousing, QC analytical laboratory

Not covered on this page: Transport and distribution, covered under good distribution practice; Cryogenic storage of cells in liquid nitrogen.

MAJOR COMPONENTS

  • Insulated chamber or room. Holds the load with defined shelving and airflow.
  • Refrigeration and heating. Holds temperature; humidifier and dehumidifier for humidity-controlled chambers.
  • Control probe. Drives the controller; not necessarily at the worst-case location.
  • Independent monitoring probe and system. Records conditions and alarms independently of the controller.
  • Door switch and backup power. Limit excursions from door openings and power loss.

Connects to: Environmental or building monitoring system; Stability study management; Warehouse and inventory systems; Emergency response and backup storage.

INSTRUMENTS AND WHAT THEY MEASURE

  • Temperature sensor (RTD or thermocouple). Control and monitoring temperature. Watch for: sensor not at the location that matters (cold spot, load centre).
  • Humidity sensor. Humidity in humidity-controlled chambers. Watch for: condensation or peroxide exposure shifting the sensor.
  • Temperature and humidity mapping loggers. Distribution across the space, empty and loaded, in qualification. Watch for: loggers placed where it is easy rather than at corners, doors and near coils.

PARAMETERS

SETPOINT · MEASURED · LIMIT · ATTRIBUTE
  • Temperature (limit, °C). The range required by the product’s storage conditions or the stability protocol, applied at the worst-case location the site found in mapping.
  • Relative humidity (limit, % RH). The range defined in the site’s stability protocol, where humidity is controlled.
  • Alarm limit and delay (setpoint, engineering units or s (delay)). Inside the storage range, with delays justified so door openings do not mask a real excursion.

SPEQ publishes no universal alarm or acceptance values: they depend on the process, the product and the site’s own qualification.

CONTROLS, ALARMS AND INTERLOCKS

  • Control loop: Temperature and humidity control. Holds setpoints.
  • Alarm: Excursion alarm with out-of-hours call-out. Warns so a response happens before stored material is affected.
  • Procedure: Excursion assessment. Impact on stored material assessed and documented after every excursion.
  • Procedure: Periodic requalification. Mapping repeated after changes and at a frequency the site justifies.

An alarm asks for a response; an interlock stops an action; a safety function protects people and has its own risk basis.

HOW IT FAILS

  • Monitoring probe at a benign location. Excursions at corners or near the door go unseen.
  • Overloading or blocked airflow. Hot or cold spots outside the mapped range.
  • Alarm not answered out of hours. A long excursion before anyone acts.
  • Power or compressor failure. Loss of conditions for the whole load.
  • Humidifier fault. Stability samples held outside protocol conditions.

ASSURANCE ACROSS THE LIFECYCLE

EXAMPLES, NOT A MANDATORY LIST
  • Qualification (Annex 15; Chapter 6, 6.29). Temperature and humidity mapping, empty and loaded; Door-open and power-failure recovery studies; Alarm function tests.
  • Routine. Continuous monitoring records; Alarm and excursion records with impact assessments.
  • Calibration and maintenance. Calibrated control and monitoring probes (Chapter 3, 3.41); Preventive maintenance.
  • Change. Re-mapping after relocation, repair or load pattern change.

WHAT THIS DOES NOT PROVE

  • A controller display in range does not prove the whole chamber is in range. Instead: Mapping results and a monitoring probe at the worst-case location.
  • Mapping when new does not prove the chamber still performs. Instead: Requalification and monitoring trends.
  • An excursion that self-corrected does not prove the stored material was unaffected. Instead: An impact assessment against the material’s stability data.

SOURCES

3 CITED
  • EU GMP Part I, Chapter 3: Premises and Equipment — European Commission (EudraLex Volume 4), 13 August 2014, in operation 1 March 2015. European Union; in force. Supports: Special storage conditions provided, checked and monitored (3.19); calibration of measuring equipment (3.41).
  • EU GMP Part I, Chapter 6: Quality Control — European Commission (EudraLex Volume 4), 28 March 2014, in operation 1 October 2014. European Union; in force. Supports: On-going stability programme (6.26–6.28); stability chambers qualified and maintained under Chapter 3 and Annex 15 (6.29).
  • EU GMP Annex 15: Qualification and Validation — European Commission (EudraLex Volume 4), 2015. EU / EEA; in force. Supports: Qualification approach.

HOW THIS PAGE WAS CHECKED

Researched and checked against the sources above by SPEQ (Claude) on 2026-10-21. It has not been reviewed by an independent engineer. It teaches how this kind of equipment works and is assured; it does not say any site, asset or process is qualified, validated or compliant.

  • ICH stability storage conditions are not stated: the status of the ICH Q1 revision was not checked here.