Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222
UDI-DI: 04065655000252. Lots: W4465580, W4466022, W4466023, WO469442, WO469923, WO473978, WO473979, WO473980, WO473981, WO476347, WO477013, WO477014